¹û¶³Ó°Ôº

XClose

¹û¶³Ó°Ôº Chemistry NMR Instruments

Home
Menu

Bruker Avance III 400

Check the queue on theÌýÌýinstrument Ìý (User ID: supervisor's initials)

A multinuclearÌý109Ag-31P(19F)-1H probe with z-gradients and automatic tuning/matching is used on this spectrometer.

The sensitivity specifications are:Ìý

1HÌý 350 : 1 (noise [200 Hz], 0.1% Ethylbenzene in CDCl3);
13C 260Ìý: 1 (noise [40 ppm], ASTM, 40% dioxane, 60% C6D6);
15NÌý 34 : 1 (noise [2 ppm], 90% formamide in DMSO-d6);
Ìý31P 217 : 1 (noise [5 ppm], 0.0485 M Triphenyl Phosphate (TPP) in acetone-d6).

There is .

Sample Case 400MHz instrument